Thai medical device labeling requirements: labels and IFU under the 2025 notification

Thai medical device labeling requirements: the short answer
Thai medical device labeling requirements for devices manufactured or imported for sale in Thailand are set by the Ministry of Public Health notification on rules, procedures and conditions for labels and instructions for use of medical devices B.E. 2568 (2025). It was published in the Royal Gazette on 22 December 2025 and repealed the 2020 notification on the same subject. A summary prepared by the Thai FDA states that the new notification took effect on 20 June 2026.
- Language Devices used by lay people need a label and instructions for use in Thai. Devices for healthcare professionals may use Thai or English.
- On the label At least 8 items, such as the product name, the manufacturer or importer, the licence number, the lot and the manufacture or expiry date.
- In the instructions for use Directions, contact details, storage and warnings, on paper or through a QR code.
- Imports The full label must be in place before sale and within 120 days of release at the checkpoint, and labels made under the 2020 rules may be used for up to 2 more years.
If you are not yet sure whether your product is a medical device or which risk class it falls into, start with what a medical device is in Thailand. The whole import process is covered in importing medical devices into Thailand, and small items such as test kits in importing small medical supplies.
Label versus instructions for use: what is the difference
The two terms are often used loosely, but the notification treats them as separate parts, each with its own list of mandatory items. The Medical Device Control Division guideline on device labelling (Non-IVD G-05) defines them as follows.
- Label: any text shown on the medical device, its container or its packaging.
- Instructions for use (IFU): paper or any other material carrying text about the device, inserted in or packed with its container or packaging, including the user manual.
Clause 6 (3) of the 2025 notification requires the label to be on the device or its container, in an easily visible place and clearly legible. Clause 6 (4) requires the instructions for use, including the user manual, on paper or in electronic form, inserted in or packed with the container or packaging.
First decide who will use the device
The language depends on the user. Clause 4 separates home use medical devices, intended for lay users outside a healthcare setting such as at home, including devices intended for use both inside and outside healthcare facilities, from professional use medical devices, intended only for healthcare professionals or for use under their supervision.
In practice, start from the intended purpose in your registration file. If the product is meant for both hospital and home use, the definition places it in the home use group, so plan for a Thai label from the start.

Who must label: manufacturers, importers and sellers
Section 44 of the Medical Device Act, as amended by the Medical Device Act (No. 2) B.E. 2562 (2019), requires establishment registrants, licensees, detailed notifiers and listing holders who manufacture or import medical devices to provide a label and instructions for use that show no false or exaggerated statements (paragraph one), following the rules set by the Minister (paragraph two). Sellers must keep the label and instructions for use provided by the manufacturer or importer (paragraph three).
Clause 5 of the 2025 notification adds that the text must be clearly visible, not false or exaggerated, not boastful about performance, and not misleading on material points. Failure to comply with Section 44 is punishable under Section 103, and failure to show warnings under Section 45 is punishable under Section 104.
Where manufacturer and importer labels differ
The clearest difference is the name and address. Clause 6 (3) (d) requires the name and address of the manufacturing site or the import site, as the case may be. An importer must also show the manufacturing site, city and country. If the manufacturer's site and city are not shown, it shows the product owner, its city and country, and the country of manufacture instead. An imported device's label therefore names both the Thai importer and the foreign source.
In practice, importers usually keep the manufacturer's label and add a Thai sticker. Agree with the product owner who designs, prints and checks it, because the Section 44 duty rests with the Thai holder of the authorisation. For the duties of shops and distributors, see medical device distributors in Thailand.
Which notification applies: the 2025 text replacing the 2020 text, and the transition
The 2020 notification was published in the Royal Gazette on 5 November 2020 and took effect 360 days later. The 2025 notification was published on 22 December 2025, took effect 180 days after publication (clause 1) and repealed the 2020 text (clause 2). The Thai FDA summary prepared in May 2026 gives the effective date as 20 June 2026.
Transitional rule: old labels may be used for 2 more years
Clause 12 provides that labels or instructions for use made under the 2020 notification that existed before the new notification came into force may continue to be used for no more than two years from the date it took effect.
In practice, existing labels from before 20 June 2026 should be replaced by June 2028. New artwork should follow the 2025 notification from the start, and keep evidence of when old labels were printed in case you need to show they existed before the effective date.
The notification does not apply to devices with their own specific labelling notification, devices exempted under Section 27, or devices made or imported for export (clause 3). Questions of interpretation are decided by the FDA Secretary-General (clause 11).
Two documents help in daily work: the G-05 guideline, with its table of headings and sample labels, and the Thai FDA summary of May 2026. In practice, note that G-05 follows the structure of the 2020 notification, so use it for format and check the mandatory items against the 2025 text every time.
- 2020 notification publishedIn force after 360 days (clause 10)
- 2025 notification publishedVolume 142, Special Part 397 Ngor, pages 19-24 · repeals the 2020 text (clause 2)
- 2025 notification takes effect180 days after publication (clause 1) · date as given in the Thai FDA summary
- Existing 2020-style labels may still be usedCounted from the effective date (clause 12)
Source: Ministry of Public Health notifications on labels and instructions for use of medical devices B.E. 2563 and B.E. 2568 · Thai FDA summary, May 2026
Thai-language medical device labels: the items that must be on the label
Clause 6 (1) requires home use devices to have a label and instructions for use in clearly legible Thai. Clause 6 (2) allows professional use devices to use clearly legible Thai or English. Both may add other languages, provided the other text does not contradict the Thai or English text.
The 8 label items under clause 6 (3)
- (a) Product name
- (b) Intended purpose or indication; if it cannot be stated on the label, it goes in the instructions for use
- (c) Number, packed quantity or fill capacity, as the case may be
- (d) Name and address of the manufacturing or import site (importers also show the source, as above)
- (e) Licence, detailed notification or listing number
- (f) Lot, batch or serial number
- (g) Month and year of manufacture, with a 4-digit year, for devices without a set shelf life
- (h) Month and year of expiry with a 4-digit year, or the shelf life counted from manufacture, for devices with a set shelf life
Items (f), (g) and (h) may be shown in English, English abbreviations or symbols. If the shelf life is stated from the date of manufacture, the month and year of manufacture must also be shown with a 4-digit year. For example, a label that says "shelf life 3 years from manufacture" must also show the manufacture month, such as 2026-05, so users can work out the expiry date.
Extra detail recommended by the G-05 guideline
- The product name should match the certificate and, for imports, the foreign manufacturer's label.
- Show only the 12-digit certificate number inside the Thai FDA frame, with no caption.
- For home use devices, the manufacturer or importer heading and the date headings need Thai captions.
- For professional use devices, if English abbreviations or symbols replace text, explain them on the label or in the instructions for use.
In practice, the Thai caption advice comes from guidance written with the 2020 notification. If you use symbols on a home use device, explain them in Thai in the instructions for use. The notification has no separate "model" item, so the model name is usually written as part of the product name so that it matches the certificate.
Source: Ministry of Public Health notification B.E. 2568, clause 6 · notification B.E. 2563, clauses 3-4 · Non-IVD G-05 guideline
Instructions for use and electronic IFU
Clause 6 (4) requires the instructions for use, including the user manual, on paper or in electronic form, such as a CD-ROM, QR code or website, inserted in or packed with the container or packaging. At least 9 items are required.
- (a) Product name
- (b) Necessary details such as key components, operating principle, category and type
- (c) Intended purpose
- (d) Indications
- (e) Directions or instructions for use, unless the device is already self-evident in use
- (f) Name, address and telephone number or other contact channel for further information or complaints
- (g) Storage conditions
- (h) Warnings, contraindications or precautions for safe use (if any)
- (i) Month and year of issue or latest revision, or the current document control code
Exemptions that save work
- If the contact details (f) are on the label, they need not be repeated in the instructions for use.
- If items (a) to (h) are all on the label, no separate instructions for use are needed.
The 2020 notification listed directions, contact details, storage and warnings among the label items, and required at least English instructions for use for professional devices. The 2025 notification moves these items to the instructions for use and allows Thai or English.
In practice, if a QR code replaces the paper manual, keep the link working for the product's life on the market, make it open on a phone without registration, provide a Thai version for home use devices, and make the version code (i) match the linked file. Keep copies of every version so you can show which lot used which manual.
Small devices, software, reusable surgical instruments and accessories
Small devices or small packaging
The 2025 notification has no separate exemption for small devices or small containers. What it does allow is placing the label on the device or on its container, moving the intended purpose to the instructions for use when it cannot fit on the label, and using abbreviations or symbols for the lot and dates.
In practice, for very small items, put the full label on the container of the smallest sales unit and explain symbols in the instructions for use. If that still cannot work, ask the Medical Device Control Division before printing, since clause 11 makes the Secretary-General the decision-maker.
Software and apps (clause 7)
Software without a physical form shows the information electronically within the software, in Thai or English, with at least 8 items, including the version number and a Unique Device Identification (UDI) for Class 4, 3 or 2 devices (if any). The Thai FDA summary adds that a screenshot showing the registered name and version may be used, and that software using AI to predict, screen, assess, analyse or diagnose should show its sensitivity and specificity.
Reusable surgical or dental instruments (clause 8)
These need Thai or English labelling with at least 5 items: product name, manufacturing or import site, lot or serial number, the manufacture date where there is no shelf life, and the expiry date or shelf life from manufacture where there is one. The 2020 notification exempted reusable surgical instruments from the general lists but required minimum labelling; the new text adds dental instruments.
Accessories (clause 9)
Accessories use the same 5 items plus the authorisation number. The Thai FDA summary explains that an accessory registered separately needs its own label, while one registered and packed together with the main device may share the main device's label. Risk classes 1 to 4 are explained in what a medical device is in Thailand.
Medical device warnings: contraindications, precautions and product-specific lists
Warnings work at two levels: general warnings, and lists set by the Minister that require certain devices to show specific statements.
Level 1: general warnings in the instructions for use
Clause 6 (4) (h) of the 2025 notification requires warnings, contraindications or precautions for safe use (if any). The G-05 guideline explains the three terms.
- Warnings: hazards the user must know before use.
- Contraindications: diseases, conditions or patient groups for whom the device should not be used, because the risk clearly outweighs the benefit.
- Precautions: special care needed for safe use, including misuse.
Level 2: devices on the warning list
Section 45, as amended by the 2019 Act, requires manufacturers and importers of devices designated under Section 6 (13) to show shelf life, warnings, contraindications or precautions on the label or in the instructions for use. The devices are listed in the Ministry of Public Health notification on medical devices required to show shelf life, warnings, contraindications or precautions on the label or instructions for use, which has been amended twice.
- 2020 notification (Royal Gazette 28 April 2020) Item 1 covered SARS-CoV-2 antibody or antigen test kits, which had to show "not for self-testing by the general public" and "not for sale to the general public" in red letters.
- No. 2 of 2020 (Royal Gazette 13 August 2020) Added item 2, surgical or isolation gowns, and item 3, coveralls. Their labels must show, in Thai or English, warnings, shelf life, and a single-use statement or the number of permitted reuses in red letters.
- No. 3 of 2024 (Royal Gazette 6 September 2024) Repealed item 1, the COVID test kits, effective the day after publication.
Failure to comply with Section 45 is punishable under Section 104. In practice, check the law library before printing each label version, because the list can change again, and translate the manufacturer's warnings faithfully: do not drop them to make the label look cleaner, and do not add statements the manufacturer has not confirmed.
Label checks at the import checkpoint: minimum data and supporting documents
Clause 10 of the 2025 notification requires the importer to present a label in Thai or English at the medical device checkpoint for inspection, with at least 5 items.
- Product name
- Manufacturing site, city and country, or product owner, city and country plus the country of manufacture
- Lot, batch or serial number
- Month and year of manufacture, where there is no shelf life
- Month and year of expiry, or shelf life from manufacture, where there is a shelf life
Items 3 to 5 may use English, abbreviations or symbols. If the manufacturer's name and city, or the dates, cannot be shown on the label, the importer may present supporting documents to the officer instead. The label and instructions for use must then be completed before sale, within 120 days of release at the checkpoint (the 2020 text allowed 180 days).
The Food and Drug Checkpoint Division guideline
The checkpoint guideline on inspecting medical device labels at import, written under the 2020 notification, contains points still useful for planning.
- The minimum text may be on a leaflet, insert or manual packed with the product, not necessarily on the label itself.
- A label in a foreign language other than English needs a certified Thai or English translation.
- If text is incomplete, bring a letter or evidence from a government agency or the manufacturer confirming the missing facts.
- If the label has no text at all, a first case brings a warning and a request for documents confirming the product matches the authorisation; if that is not possible, the goods are sent back to the country of origin.
In practice, ask the manufacturer to confirm before shipping that the source label carries the minimum data, and build the 120 days into your stickering plan when you place the order. Foreign certificates are discussed in CFS vs EC certificate for medical devices.
Checklist before you file, and common mistakes in practice
Before artwork goes into the registration file or to the printer, run through the checklist below. The Medical Device Control Division also provides a template comparing the old and proposed label and instructions for use, used when applying to change a label.
Common mistakes in practice
- Using the old list New labels follow the G-05 table or an old label without checking against the 2025 text.
- Product names do not match The label, certificate and manufacturer's label write the model differently, which needs explaining both at filing and at the checkpoint.
- Year not in 4 digits The lot printer is still set to a 2-digit year.
- English-only home use devices Common for products sold both to hospitals and in retail.
- Wrong type of number A cosmetic notification number left on a product that has become a medical device.
- Broken QR code The link changed after a website move, or the page is English only.
- Exaggerated text Marketing copy taken from a foreign box, which conflicts with Section 44 and clause 5.
- Forgetting the clock after release The Thai label must be finished before sale and within 120 days.
- ✓User group decided: home use or professional use2025 notification, clauses 4 and 6 (1)-(2)
- ✓All 8 label items under clause 6 (3) (a)-(h)Years in 4 digits
- ✓Product name and certificate number match the certificateG-05 guideline
- ✓IFU has all 9 items, or the label is complete enough for the exemptionClause 6 (4)
- ✓QR code or website for the manual works and shows a version codeClause 6 (4) (i)
- ✓Checked whether the product is on the specific warning listSection 45 · 2020 notification and amendments
- ✓Software, reusable instruments and accessories use clauses 7-9Including UDI for Class 2-4 software (if any)
- ✓Imports: checkpoint minimum data and a plan to finish the Thai label within 120 daysClause 10
- ✓No false, exaggerated or boastful statementsSection 44 paragraph one · clause 5
Source: Medical Device Act · Ministry of Public Health notification B.E. 2568 · Non-IVD G-05 guideline
Frequently asked questions
Must a medical device label in Thailand be in Thai?
It depends on the user. Home use devices need Thai; professional use devices may use Thai or English, under clause 6 (1) and (2) of the 2025 notification.
How long can labels made under the 2020 notification be used?
Clause 12 allows them for no more than two years from the date the 2025 notification took effect, which is 20 June 2026. In practice, plan to finish the change by June 2028.
Can a QR code replace the paper manual?
Yes. Clause 6 (4) allows electronic instructions for use, such as a QR code or website, but the 8 label items must still appear on the label.
Can a Thai label be added after import?
Yes. At the checkpoint the minimum data must be in Thai or English, and the full label must be completed before sale, within 120 days of release, under clause 10.
What if a device is too small for all the information?
The 2025 notification has no separate small-device exemption, but the label may go on the container, the intended purpose may move to the instructions for use, and symbols may be used for the lot and dates. If that still does not work, ask the Medical Device Control Division.
Which medical devices need warnings in red letters?
Under amendment No. 2 of 2020, medical gowns and coveralls must show a single-use statement or the number of permitted reuses in red letters. The COVID test kit entry was repealed by No. 3 of 2024.
Sources
- Medical Device Act B.E. 2551 (2008) as amended (Council of State consolidated text) · Open source · Search our archive
- Medical Device Act (No. 2) B.E. 2562 (2019), Sections 29, 30 and 50 amending Sections 44, 45, 103 and 104 · Open source · Search our archive
- Ministry of Public Health notification on labels and instructions for use of medical devices B.E. 2568 (2025), clauses 1-12 · Open source · Search our archive
- Ministry of Public Health notification on labels and instructions for use of medical devices B.E. 2563 (2020), repealed · Open source · Search our archive
- Ministry of Public Health notification on medical devices required to show shelf life, warnings, contraindications or precautions B.E. 2563 (2020) · Open source · Search our archive
- Same notification, No. 2 B.E. 2563 (2020): medical gowns and coveralls · Open source · Search our archive
- Same notification, No. 3 B.E. 2567 (2024) · Open source · Search our archive
- Guideline on labels and instructions for use of medical devices (Non-IVD G-05), Medical Device Control Division · Open source · Search our archive
- Requirements and standards for medical device labels and IFU (Thai FDA summary, May 2026) · Open source · Search our archive
- Guideline for inspecting medical device labels at the import checkpoint (Food and Drug Checkpoint Division) · Open source · Search our archive
- Template comparing current and proposed medical device labels and instructions for use · Open source · Search our archive
- Ministry of Public Health notification on medical device classification by risk B.E. 2562 (2019) · Open source · Search our archive
This article is for general information. It is not a Thai FDA ruling or legal advice, and rules change — check the latest notification or ask an expert before filing or producing. Spotted an error? Tell us and we will correct it