Thai Medical Device Act B.E. 2551 Explained: Definition, Structure, Licences, Labels, Advertising and the 2019 Amendment

What is the Thai Medical Device Act B.E. 2551, and who does it apply to: manufacturers, importers, sellers and healthcare facilities
The Thai Medical Device Act B.E. 2551 (2008) is the parent law that governs medical devices in Thailand. It says what counts as a medical device, who must register an establishment before manufacturing or importing, whether each group of products needs a licence, a detailed notification or a listing, what operators must do once products are on the market, and that advertising needs a permit first. Topic-level detail, such as risk-based grouping, quality systems, labels and responsible persons, sits in ministerial regulations and notifications issued under the Act.
- In force: from the day after its publication in the Royal Gazette (Section 2). It repealed the Medical Device Act B.E. 2531 (1988) (Section 3).
- Amended by No. 2: the Medical Device Act (No. 2) B.E. 2562 (2019) replaced the definition of a medical device, split control of manufacture and import into 3 risk-based groups, added listing, and added a part on the medical device permit review process (details in section 9).
- Register the establishment first: anyone who intends to manufacture or import medical devices must register the establishment with the permit grantor (Section 15), then apply for a licence, file a detailed notification or file a listing according to the group set by notification (Sections 17 and 19).
- Products that may not be sold: no one may manufacture, import or sell counterfeit, substandard, deteriorated or unsafe medical devices (Section 46), and no one may sell a medical device that has no licence, detailed-notification certificate or listing certificate (Section 46/1).
Who it applies to
The duties hang on the definitions in Section 4. To manufacture means to make, assemble, fabricate, repack, pack together, modify, process, adapt or sterilise. To sell includes distributing, giving away, exchanging, lending, renting out or hire-purchase for commercial benefit, and having for sale, so renting out equipment commercially is covered too. The Act also places duties on healthcare facilities in some cases, such as the patient registry in Section 42 and the liability of anyone who uses a medical device on another person in Section 78.
This article is an overview. The statutory text is quoted from the consolidated file in this website's archive, whose footnote records publication in the Royal Gazette, Volume 125, Part 43 Kor, page 25, dated 5 March 2008 (B.E. 2551), and which marks the sections amended by No. 2. Before citing it in an official document, check the current version in the Royal Gazette.
The definition of a medical device in Section 4: software, test reagents, accessories and the no-pharmacological-action rule
Every duty starts from the definitions in Section 4, which the No. 2 Act of 2019 repealed and replaced with new wording. This section summarises the statutory text only; risk grouping with product examples is covered in what is a medical device in Thailand.
The 3 limbs of the definition
- an instrument, apparatus, machine, implant, reagent used for testing inside or outside a laboratory, product, software or other article that the manufacturer or product owner intends specifically for use on humans or animals for one of the 8 purposes in items (a) to (h), such as diagnosing, preventing, monitoring or alleviating disease, supporting or sustaining life, and providing information from examining specimens taken from the body for diagnosis;
- an accessory for use with a medical device under limb 1; and
- an instrument, apparatus, machine, product or other article that the Minister declares by notification to be a medical device.
The closing condition: the main effect must not come from drug action
The definition ends by requiring that the intended effect of an article under limb 1 in the human or animal body must not be achieved principally by pharmacological, immunological or metabolic means. In practice, if a product's main effect comes from the action of a substance, compare it with drug law first; if it is used externally for cleanliness or beauty, compare it with the Thai Cosmetics Act B.E. 2558; and for cosmetics in containers that push substances into the skin, read microneedles in cosmetics.
Accessories, software and other definitions worth knowing
- Accessory (added by No. 2): an article, appliance or product that the manufacturer or product owner intends specifically for use with a medical device, to help it or enable it to work for its intended purpose.
- Software and test reagents appear directly in limb 1, so software or an app can be a medical device if the manufacturer intends it for purposes (a) to (h).
- Label and instructions: a label is any statement shown on the device, its container or its package; the accompanying document includes the instructions for use.
- Advertising includes sales promotion, meaning giving information, persuading or doing anything to induce a sale.
In practice, what decides the category is the purpose the manufacturer intends, so the wording on labels, manuals and sales material carries great weight. Start with the product classification tool before ordering stock or printing packaging.

Medical Device Act B.E. 2551 structure: 12 chapters, with Section 129 as the last section
Counting from the consolidated text, the Medical Device Act B.E. 2551 ends at Section 129 (archive file, page 39). It is divided into an opening part (Sections 1 to 6), 12 chapters and transitional provisions (Sections 124 to 129). The No. 2 Act of 2019 added slash-numbered sections such as 30/1, 31/1, 35/1 to 35/7, 46/1 and 70/1 and repealed Section 33, so the number of sections in force is not the same as the last section number.
- Opening part: title, entry into force, repeal of the old law, definitions and the power to issue ministerial regulations and notifications (Sections 1 to 6)
- Chapter 1: the Medical Device Committee (Sections 7 to 14)
- Chapter 2: establishment registration, applying for and granting permits, detailed notification and listing (Sections 15 to 35), including Part 1 on the medical device permit review process (Sections 35/1 to 35/7)
- Chapter 3: closing and transferring a business (Sections 36 to 39)
- Chapter 4: duties of establishment registrants, licensees, detailed notifiers, listing holders and sellers (Sections 40 to 43)
- Chapter 5: labels and accompanying documents (Sections 44 and 45)
- Chapter 6: control of medical devices (Sections 46 to 55)
- Chapter 7: advertising (Sections 56 to 60)
- Chapter 8: competent officials (Sections 61 to 66)
- Chapter 9: suspension and revocation of establishment registrations, licences and detailed-notification certificates, and cancellation of listing certificates (Sections 67 to 73)
- Chapter 10: appeals (Sections 74 to 76)
- Chapter 11: civil liability (Sections 77 to 80)
- Chapter 12: penalties (Sections 81 to 123)
The titles of Chapters 2, 4 and 9 above are as amended by No. 2. The Thai FDA's unofficial English translation calls a detailed notification a "specification notification" and a listing a "notification"; this article uses licence, detailed notification and listing, as the other medical device articles on this site do. The Medical Device Committee is chaired by the Permanent Secretary of Public Health, and the Director of the Medical Device Control Division of the Thai FDA sits on it as member and assistant secretary (Section 7). The Committee advises the Minister on issuing notifications (Section 11). In practice, when you open a notification, check whether it was issued by the Ministry of Public Health, the Committee or the Thai FDA, and under which section.
Source: Medical Device Act B.E. 2551, consolidated text (archive file, pages 1–39) · right-hand column is practical advice
How Thai medical device law works: the Act sets the frame, notifications under Section 6 fill in the detail
The Act sets a broad frame and delegates the detail to secondary legislation, so most of what an RA team checks day to day is in ministerial regulations and notifications, not in the Act itself.
Section 5 gives the Minister of Public Health the power to appoint competent officials and to issue ministerial regulations and notifications, which take effect once published in the Royal Gazette. Section 6 lets the Minister, on the Committee's advice, prescribe a range of matters; it now runs from item (1) to item (22), with item (2) repealed. The items operators use most are:
- Risk groups (1): groups that manufacturers or importers must license (a), file a detailed notification for (b) or list (c)
- Sales (3), (9), (10): groups that sellers need a licence for, and devices that may be sold only to consumers with a prescription, or only to healthcare facilities or practitioners
- Standards and systems (4) to (7): device standards, quality systems, containers, and responsible persons for manufacture, import or sale
- Bans and labels (11) to (13): devices that may not be produced, imported or sold, devices barred from direct selling, and devices that must show shelf life, warnings or precautions on the label
- Other: technology assessment (8), patient registries, clinical research, transport and destruction (14) to (16), import checkpoints (17), exemptions (18), and the costs of the permit review process that No. 2 added as (19) to (22)
Examples of notifications in the archive
- Risk classes: the Notification on classifying medical devices by risk B.E. 2562 (2019) sets Classes 1 to 4 for both in vitro diagnostic and other devices; Class 4 needs a licence, Classes 2 and 3 a detailed notification, and Class 1 a listing.
- Quality systems and responsible persons: the Notification on manufacturing quality systems B.E. 2566 (2023), which offers ISO 13485:2016 as one option, the Notification on import and sale quality systems B.E. 2566 (2023), and the 2024 Notification on responsible persons, which requires at least one responsible person.
- Sales: the 2020 Notification on devices that may be sold only to healthcare facilities or practitioners, and the 2024 Notification on devices barred from direct selling, which covers devices not used by the general public and does not ban every device.
- Labels: the Notification on labels and accompanying documents B.E. 2568 (2025), which repealed the 2020 version.
In practice, before citing a notification, check that it has not been repealed or amended. How to search is explained in how to find Thai FDA notifications, and the notifications themselves can be searched on this website’s notification search page.
Establishment registration, then a licence, detailed notification or listing: Sections 15 to 35
Step one: establishment registration, Sections 15 and 16
Section 15 requires anyone who intends to manufacture or import medical devices to register the establishment with the permit grantor. Section 16 sets the requirements, such as being at least twenty years old, residing in Thailand, and having manufacturing or import premises with the equipment the Minister prescribes. The premises criteria are in the 2009 ministerial regulations on registering manufacturing and import establishments.
Step two: file according to the risk group
- Licence (Section 17): devices under Section 6(1)(a) need a licence application, and may be manufactured or imported once the licence is issued, under the 2020 ministerial regulation on licensing.
- Detailed notification or listing (Section 19): devices under Section 6(1)(b) or (c) need a detailed notification or a listing, and may be manufactured or imported once the detailed-notification certificate or listing certificate is issued, under the 2020 ministerial regulation for each route.
- Existing operators (Sections 18 and 20): when a new group notification takes effect, they must apply within the period that notification states, and may continue until an order refusing the application.
For the documents to obtain from a foreign product owner, read importing medical devices into Thailand, CFS vs EC certificate for medical devices and importing small medical supplies.
Sales licence: Sections 24 to 26
Section 24 requires anyone who will sell medical devices under Section 6(3) to obtain a licence first. A manufacturer or importer under Section 17 or 19 is deemed to hold a sales licence for the devices it manufactures or imports, without applying for one. Section 26 requires a sales applicant to have a sales quality system under Section 6(5) and a responsible person for sales where the device falls under Section 6(7). Read more in medical device distributor in Thailand.
Validity, exemptions and export
- Validity (Section 29): establishment registrations, licences and detailed-notification certificates are valid until 31 December of the fifth year from the year of issue; a sales licence is valid until 31 December of the year of issue; and a listing certificate is valid for five years from the date shown on it.
- Renewal and amendment (Sections 30 to 31/1): renew before expiry; within one month after expiry a renewal with a request for leniency is possible, but it does not excuse the offence under Section 91. Changes require an application, and a listing holder must report when listed information changes.
- Exemptions (Section 27): Sections 15, 17, 19 and 24 do not apply in certain cases, such as state agencies and the Thai Red Cross, a practitioner making a device for his or her own individual patient, and manufacture or import in quantities needed for personal use, samples, exhibitions or research, subject to the criteria set by notification.
- Export (Section 34): devices may be made for export to the quality, standard or labelling the buyer specifies, but such devices may not be sold in Thailand.
Source: Medical Device Act B.E. 2551, consolidated text, pages 8–12 · MOPH Notification on classifying medical devices by risk B.E. 2562, page 2
Operator duties, labels and instructions for use: Chapters 4 and 5
Premises and core duties: Sections 40 and 41
Section 40 prohibits establishment registrants, licensees, detailed notifiers and listing holders from manufacturing, importing, selling or storing medical devices outside the premises named in their documents, except for permitted temporary storage, direct sales to healthcare practitioners and assembly for installation under the prescribed criteria. Section 41 sets 9 duties, summarised as follows:
- Quality system and responsible person: follow the quality system under Section 6(5) and appoint a responsible person under Section 6(7) (items 1 and 2)
- Records and reports: keep records of manufacture, import or sale and report to the permit grantor (item 3), under the 2025 notification
- Adverse events and complaints: report malfunctions, adverse events and field safety corrective actions, whether they occur in Thailand or abroad, and run a complaint-handling system (items 4 and 5)
- Signs and documents: premises signs, a sign with the responsible person's name and qualifications, display of the registration or sales licence, and technical documents supporting quality, standard, performance and safety (items 6 to 9)
Section 42 requires a patient registry for devices under Section 6(14), and Section 43 allows devices under Section 6(9) or (10) to be sold only to consumers with a prescription, or only to healthcare facilities or practitioners.
Labels and accompanying documents: Chapter 5
Section 44 requires establishment registrants, licensees, detailed notifiers and listing holders who manufacture or import medical devices to provide labels and accompanying documents that do not contain false or exaggerated statements, following the criteria the Minister prescribes. Its third paragraph requires sellers to keep the labels and documents that the manufacturer or importer provided. Section 45 requires devices covered by a notification under Section 6(13) to show their shelf life, warnings, contraindications or precautions on the label or accompanying document.
The 2025 label notification requires Thai-language labels and documents for devices used by the general public, allows Thai or English for devices used by professionals, and has specific criteria for software without a physical form. The required items and the transition from the 2020 version are summarised in Thai medical device labeling requirements. In practice, check the draft with the label checker before printing or before the goods leave the port.
Medical devices that may not be produced, imported or sold: counterfeit, substandard, deteriorated and unsafe devices
Section 46 prohibits anyone from manufacturing, importing or selling 6 kinds of medical device: counterfeit, substandard, deteriorated, unsafe, manufactured or imported differently from what was licensed, notified or listed, and devices whose licence or detailed-notification certificate has been revoked or whose listing certificate has been cancelled. Section 46/1, added by No. 2, prohibits anyone from selling a medical device that has no licence, detailed-notification certificate or listing certificate.
Counterfeit devices: Section 47
A counterfeit device is one that is imitated or copied in whole or in part; one that misleads about its name, components, quality, quantity, date of manufacture, expiry date, manufacturer, place of manufacture, importer, quality mark or trademark; or one that claims to be licensed, notified or listed when that is untrue.
Substandard and deteriorated devices: Sections 48 and 49
A substandard device is one whose quality or standard does not match what was licensed, notified or listed, or that does not meet the standard under Section 6(4) or the container standard under Section 6(6), except devices made for export under Section 34. A deteriorated device is one that has turned into a substandard device or has passed the shelf life shown on it. In practice, sellers should check the shelf life of their stock regularly.
Unsafe devices: Section 50
- single-use and already used
- manufactured or stored unhygienically
- containing foreign matter or something likely to harm health, or a decomposable substance that may become toxic
- of unreliable benefit
- designed or made in a way that may harm the user
- with a label or accompanying document that does not comply with Section 44 or 45 in a way that may harm the user
Powers when there is doubt or a problem: Sections 51 to 55
Imports must pass inspection at a medical device checkpoint once a notification under Section 6(17) is in force (Section 53). If a device is suspected of lacking quality or being unsafe, the Secretary-General may order evidence to be submitted and suspend manufacture, import or sale temporarily (Section 54). Where a problem is found, the Secretary-General may order corrections, stop manufacture, import or sale, publish the findings, and recall the device from the market, with the manufacturer, importer, seller or possessor bearing the cost (Section 55).
Medical device advertising, competent officials, suspension and appeals: Sections 56 to 76
Advertising: Sections 56 to 60
Section 56 prohibits advertising devices banned under Section 6(11) or devices under Section 46. Section 57 requires a permit from the permit grantor before advertising a medical device; an advertising permit is valid for no more than three years from issue, and the permit grantor may set specific conditions and restrict the media used. Its fourth paragraph, added by No. 2, lets the Minister prescribe devices and criteria under which advertising directly to healthcare practitioners needs no permit.
The 2020 Notification on certain medical device advertising that needs no permit exempts advertising that shows only a trade name, trademark or trade logo without stating properties, benefits, quality, standard, composition or origin, for devices whose manufacturer or importer has complied with Sections 15, 17 and 19. Section 59 requires that advertising must not:
- state false or exaggerated benefits, quality, quantity, standard, composition or origin
- show any person endorsing or praising the device's benefits
- offer prizes by chance in any form
- claim the device can prevent, cure or alleviate diseases or symptoms that the Minister has barred from advertising
- contain statements that mislead on material points
If advertising breaches Section 57 or 59, the permit grantor may order the wording changed, ban certain wording or stop the advertising, and may also order correct information to be published (Section 60). In practice, an online sales post that states a product's benefits falls within the definition of advertising, since it includes sales promotion, so a permit should be obtained before publishing, and endorsements or reviews praising benefits should not be used.
Competent officials: Chapter 8
Section 61 lets competent officials inspect premises for manufacture, import, sale or storage during business hours, take samples, seize or attach goods and demand documents. Operators must facilitate them (Section 63).
Suspension, revocation, cancellation and appeals: Chapters 9 and 10
The permit grantor, with the Committee's approval, may suspend or revoke establishment registrations, licences or detailed-notification certificates under Sections 67, 69 and 70, for example where a device is unsafe or its intended use has been changed into a drug or cosmetic without permission, and may cancel a listing certificate under Section 70/1, for example where the listed information does not match the facts. Anyone refused a document, or whose document is suspended, revoked or cancelled, may appeal in writing to the Minister within thirty days (Sections 74 and 75). Chapter 12, Sections 81 to 123, sets penalties by the section breached; this article does not summarise penalty rates, and anyone facing a real case should consult a legal expert.
What the Medical Device Act No. 2 B.E. 2562 changed, and civil liability in Chapter 11
The Medical Device Act (No. 2) B.E. 2562 (2019)
The No. 2 Act has 64 sections, was published in the Royal Gazette, Volume 136, Part 56 Kor, page 186, dated 30 April 2019, and took effect the day after publication (Section 2). Its changes fall into 5 groups:
- Definitions: repealed and replaced the definition of a medical device (Section 3), and added definitions of accessory (Section 4) and listing holder (Section 5).
- Risk-based system: amended Section 6(1) into 3 groups, licence, detailed notification and listing (Section 6), amended Sections 17 to 20 (Section 15), and added the five-year validity of listing certificates to Section 29 (Section 20).
- Permit review process: added Part 1, Sections 35/1 to 35/7 (Section 24), which must also give weight to promoting domestic manufacture of medical devices (Section 35/1); besides Thai FDA officials, experts, expert organisations, state agencies or private organisations in Thailand or abroad may assess technical documents, carry out testing or inspect establishments, provided they are listed by the Thai FDA (Section 35/2).
- Duties and control: renamed Chapter 4 to cover listing holders and sellers, amended Sections 40, 41 and 44, added Section 46/1, repealed Section 33, added the fourth paragraph of Section 57 and added Section 70/1.
- Transition: pending applications are treated as applications under the amended law (Section 60), existing licences and detailed-notification certificates remain valid until expiry (Section 61), and existing ministerial regulations and notifications continue so far as they are consistent (Section 62).
The note at the end of the Act explains that Thailand had entered into the ASEAN Agreement on Medical Device Directive, so the control measures were amended to match it by grouping devices by risk, adding listing, reducing export controls and making the permit review process more efficient.
Civil liability: Chapter 11
This chapter has existed since the 2008 Act, whose closing note says it added civil liability provisions. No. 2 did not amend Sections 77 to 80.
- Manufacturers, importers and sellers (Section 77) are liable for damage caused by using the medical device, unless they prove the damage resulted from force majeure, did not arise from a defect in the device, or was the injured person's own fault.
- Anyone who uses a device on another person (Section 78) is liable for resulting harm to life, body or health, unless they prove they exercised care in line with professional standards, or the harm resulted from force majeure or the injured person's own fault.
- Prescription (Section 79): a claim for damages under this chapter is barred three years after the injured person learns of the damage and of the person liable, and in any event no more than ten years after the damage occurred.
- Recourse (Section 80): a person who has paid damages may claim recourse from others who contributed to the damage within three years of paying, only for the portion beyond their own liability.
In practice, importers and distributors should keep sales records and lot numbers so products can be traced, and agree recourse with the manufacturer or product owner in the contract; clinics that use devices on clients should read Section 78 alongside the procedures in the instructions for use.
After reading the Medical Device Act, what next: a checklist by type of business
The next step is to match your business role to the sections and notifications you must follow. This list is practical advice, not the text of the law.
- Every business: first confirm that the product is a medical device under Section 4 with the product classification tool, then check whether it falls in the licence, detailed-notification or listing group and whether a product-specific notification applies.
- Importers: register an import establishment under Section 15, file by risk group under Section 17 or 19, prepare the product owner's documents with the import document checker, appoint a responsible person for import, set up the 2023 import and sale quality system, and apply Thai labels before sale. Read importing medical devices into Thailand.
- Distributors that do not import: check that every product you take on has a licence, detailed-notification certificate or listing certificate, because Section 46/1 bans selling products without one, and check whether the product is in a group that needs a sales licence under Section 6(3). Read medical device distributor in Thailand.
- Local manufacturers: register a manufacturing establishment, set up the manufacturing quality system under the 2023 notification, appoint a responsible person for manufacture, keep technical documents under Section 41(9), and see Section 34 when making to a foreign buyer's specification.
- Clinics, shops and online sellers: sell only products with documents under Section 46/1, check whether a product may be sold only on prescription or only to healthcare facilities under Section 43, obtain an advertising permit before posting benefits under Section 57, and clinics should read Section 78 on the liability of device users.
- Software and app developers: write down the intended purpose clearly and compare it with items (a) to (h) of the definition. If it fits, plan as for any other medical device: the manufacturer or importer registers an establishment and files by risk group. The 2025 label notification has criteria for software without a physical form. Read software and apps that are medical devices.
Then note the expiry date of every document under Section 29, apply before changing any listed details under Sections 31 and 31/1, and track new notifications on the notification search page. An overview of filing steps for every product group is in Thai FDA application steps. Businesses with several product groups should read this together with the Thai Cosmetics Act B.E. 2558, the Thai Food Act B.E. 2522 and the Thai herbal products law.
- ✓Confirm it is a medical deviceCompare the intended purpose with the Section 4 definition, including software and test reagents
- ✓Register the establishmentManufacturers and importers register first (Section 15)
- ✓File by risk groupLicence, detailed notification or listing (Sections 17 and 19)
- ✓Quality system and responsible personQuality system under Section 6(5) and responsible person under Section 6(7) (Section 41)
- ✓Labels, advertising and salesLabels under Section 44 · advertising permit before publishing (Section 57) · no sale of products without documents (Section 46/1)
- ✓After market entryRecords, adverse event reports, complaints, and renewal before expiry (Sections 29 to 30/1)
Source: Practical advice citing section numbers from the Medical Device Act B.E. 2551, consolidated text
Frequently asked questions
How many chapters does the Medical Device Act B.E. 2551 have, and what is the last section?
The consolidated text has an opening part, 12 chapters and transitional provisions, ending at Section 129. The 12 chapters cover the Committee, establishment registration and permits, closing and transferring a business, operator duties, labels and accompanying documents, control, advertising, competent officials, suspension and revocation, appeals, civil liability and penalties. The No. 2 Act of 2019 added slash-numbered sections such as 35/1 to 35/7 and 46/1.
Can software or an app be a medical device in Thailand?
Yes. The Section 4 definition, as amended by No. 2, lists software directly. If the manufacturer or product owner intends it for one of the purposes in items (a) to (h), such as diagnosing or monitoring disease, or providing information from examining body specimens for diagnosis, the software is a medical device.
How long are a licence, detailed-notification certificate and listing certificate valid?
Under Section 29, establishment registrations, licences and detailed-notification certificates are valid until 31 December of the fifth year from the year of issue, a sales licence until 31 December of the year of issue, and a listing certificate for five years from the date shown on it. Renewal must be filed before expiry.
Does every medical device seller need a sales licence?
No. The sales licence under Section 24 applies to devices prescribed under Section 6(3), and manufacturers and importers are deemed to hold a sales licence for what they make or import. But every seller is bound by Section 46/1, which bans selling a medical device that has no licence, detailed-notification certificate or listing certificate.
Does medical device advertising need a permit?
Section 57 requires a permit before advertising a medical device, valid for no more than three years. A 2020 notification exempts advertising that shows only a trade name, trademark or logo without stating properties or benefits, and Section 59 bans false or exaggerated statements, endorsements by individuals and prize draws.
Which text does this article cite, and is fdathailand.com a Thai FDA website?
This article cites files in the fdathailand.com archive: the consolidated B.E. 2551 Act with its amendment footnotes, and the No. 2 Act of B.E. 2562 as published in the Royal Gazette. The law may be amended later; before relying on it, check the current version in the Royal Gazette. fdathailand.com is an independent website, not a website of the Thai Food and Drug Administration.
Sources
- Medical Device Act B.E. 2551 (2008), consolidated text as amended by No. 2 B.E. 2562 (2019) · Open source · Search our archive
- Medical Device Act (No. 2) B.E. 2562 (2019) · Open source · Search our archive
- Medical Devices Act B.E. 2551 (2008) (unofficial English translation) · Open source · Search our archive
- MOPH Notification on classifying medical devices by risk B.E. 2562 (2019) · Open source · Search our archive
- MOPH Notification on groups of medical devices requiring detailed notification (No. 3) B.E. 2565 (2022) · Open source · Search our archive
- Ministerial Regulation on registering medical device manufacturing establishments B.E. 2552 (2009) · Open source · Search our archive
- Ministerial Regulation on registering medical device import establishments B.E. 2552 (2009) · Open source · Search our archive
- Ministerial Regulation on licences to manufacture or import medical devices B.E. 2563 (2020) · Open source · Search our archive
- Ministerial Regulation on detailed notification to manufacture or import medical devices B.E. 2563 (2020) · Open source · Search our archive
- Ministerial Regulation on listing to manufacture or import medical devices B.E. 2563 (2020) · Open source · Search our archive
- Ministerial Regulation on licences to sell medical devices B.E. 2555 (2012) · Open source · Search our archive
- MOPH Notification on quality systems for manufacturing medical devices B.E. 2566 (2023) · Open source · Search our archive
- MOPH Notification on quality systems for importing or selling medical devices B.E. 2566 (2023) · Open source · Search our archive
- MOPH Notification on medical devices requiring responsible persons, and their qualifications, number and duties B.E. 2567 (2024) · Open source · Search our archive
- MOPH Notification on medical devices that may be sold only to healthcare facilities or practitioners B.E. 2563 (2020) · Open source · Search our archive
- MOPH Notification on medical devices requiring a sales licence and sale to consumers with a prescription B.E. 2567 (2024) · Open source · Search our archive
- MOPH Notification on medical devices barred from direct selling or direct marketing B.E. 2567 (2024) · Open source · Search our archive
- MOPH Notification on medical devices requiring technology assessment B.E. 2563 (2020) · Open source · Search our archive
- MOPH Notification on labels and accompanying documents of medical devices B.E. 2568 (2025) · Open source · Search our archive
- MOPH Notification on records and reports of manufacture, import or sale of medical devices B.E. 2568 (2025) · Open source · Search our archive
- MOPH Notification on reporting malfunctions, adverse events and field safety corrective actions B.E. 2563 (2020) · Open source · Search our archive
- MOPH Notification on complaint channels, records and complaint-handling systems B.E. 2563 (2020) · Open source · Search our archive
- MOPH Notification on certain medical device advertising that needs no permit B.E. 2563 (2020) · Open source · Search our archive
This article is for general information. It is not a Thai FDA ruling or legal advice, and rules change — check the latest notification or ask an expert before filing or producing. Spotted an error? Tell us and we will correct it